DWY-60A Impact test low temperature meter
....................................................................................................................................................................................................................................................

Keywords: low temperature meter cryotank Impact low temperature meter Impact sample low temperature meter Nanjing low temperature meter Nanjing cryotank Nanjing Impact low temperature meter

First, the device name: Impact test low temperature meter

Two, Model: DWY-60A

Third, use and features:

Impact test temperature recommendations are based on the Company GB229-2002 "Charpy Impact test methods" requirements for low temperature means the latest developed compressor refrigeration equipment. This equipment uses cascade compressor refrigeration technology, the use of heat balance principle and circulation mixing method, to achieve automatic uniform cooling of the sample, temperature, temperature indicators fully meet the national standards required by GB-229-2002. The device is simple and convenient operation, high efficiency is the material of the best low-temperature Impact test specimen cooling, insulation equipment. But it will also be used for other low temperature testing and experimental work.

Fourth, the main technical parameters:

Temperature range: + 30 ℃ ~ -60 ℃ (room temperature ≤25 ℃)

Temperature accuracy: <± 0.5 ℃

Cooling rate: + 30 ℃ ~ 0 ℃ about 2.0 ℃ / min

0 ℃ ~ -20 ℃ about 1.5 ℃ / min

-20 ℃ ~ -40 ℃ about 1.0 ℃ / min

-40 ℃ ~ -60 ℃ about 0.7 ℃ / min

Maximum dimensions: 900 × 505 × 850mm

Effective test space: 120 × 120 × 70mm

Number of samples can be loaded:> 60 (Impact sample size: 10 × 10 × 55mm)

Digital display Timer: 1 minute to 99 hours, 1 minute resolution

Cooling medium: ethanol or other antifreeze (not included)

Stirring motor: 8W

Power supply: 220V ~ 240V, 50HZ, 3KW

Fifth, technical parameters + configuration:

1. Import dual compressor;

2. The sample basket.

3. The sample holder.

 